Structure of an Atom and Mass Spectrum
Senior 4 · Chemistry
Structure of an Atom and Mass Spectrum
Lesson 3 of 4
25 min
The last lesson used numbers such as "75.76% chlorine-35". Nobody can pick chlorine atoms out one by one and put them on a balance. Those numbers come from a mass spectrometer, an instrument that turns atoms or molecules into ions, sorts the ions by mass, and counts how many there are of each. Today mass spectrometers measure isotope ratios, identify unknown compounds, test athletes' urine for banned drugs and analyse the soil of Mars. This lesson follows a sample through a classic magnetic-sector instrument, stage by stage.
Definition
Mass spectrometer: an instrument that separates gaseous ions according to their mass-to-charge ratio and records how many ions of each ratio there are.
Mass-to-charge ratio, m/z: the mass of an ion (in u) divided by the number of charges on it. For a 1+ ion, m/z is simply its relative mass: has m/z = 35, while has m/z = .
The whole instrument is kept at a high vacuum. Otherwise the ions would collide with air molecules, which would knock them off course, slow them down or neutralise them.

An ion of mass and charge is accelerated through a potential difference , so its kinetic energy is
In a magnetic field at right angles to its motion, the magnetic force supplies the centripetal force for a circular path of radius :
Eliminating between (1) and (2):
For fixed and , the radius depends only on : . A heavier ion follows a wider curve (it is deflected less); a more highly charged ion follows a tighter curve (it is deflected more). Two different ions with the same m/z, such as and , follow exactly the same path, so the instrument measures m/z, not mass alone.
Worked example 1: which ion is deflected most?
The ions , , , and pass through the same instrument. List them from most deflected to least deflected.
Solution. Work out m/z for each: ; ; ; ; . The smaller m/z, the greater the deflection.
Answer: (1) > (2) > (4) > (10) > (12).
Worked example 2: radius of the path
With a certain magnetic field and accelerating voltage, ions follow a path of radius 10.0 cm. Without changing the settings, what is the radius of the path of ions? What must be done to bring the ions onto the detector instead?
Solution. Since and both ions are 1+:
Answer: 10.5 cm (3 s.f.). The heavier ions follow a wider arc. To make them follow the 10.0 cm path to the detector, the magnetic field must be made stronger (by a factor of 1.049), since .
Many modern instruments have no magnet. In a time-of-flight (TOF) mass spectrometer, every ion is given the same kinetic energy and then drifts down a field-free tube of length . From equation (1), lighter ions move faster and arrive first. The flight time is
The detector records the arrival times, and a computer converts each time into an m/z value. Whatever the design, the output is the same: a mass spectrum, a graph of ion abundance against m/z, which the next lesson teaches you to read.
Warning
Common mistakes. (1) Heavier ions are deflected less, not more. (2) The spectrometer separates by m/z, not by mass: a 2+ ion appears at half its mass. (3) Ions are positive because electrons are knocked off; do not write for this kind of ionisation. (4) The vacuum is there to prevent collisions, not to "pull the ions along".
Tip
Learn the five stages in order with the first letters V I A D D: Vaporise, Ionise, Accelerate, Deflect, Detect. For each stage in an exam answer, say what happens and how: "Ionise: bombarded with high-energy electrons from an electron gun, which knock out an electron to form ."