Radiation and Medical Imaging
Senior 6 · Physics
Radiation and Medical Imaging
Lesson 4 of 5
25 min
On 1 October 1971, at Atkinson Morley's Hospital in London, a woman with a suspected brain tumour lay with her head in a new machine built by the engineer Godfrey Hounsfield. An X-ray beam scanned round her head for several minutes, a computer then spent hours processing the readings, and the image showed a dark, round cyst in the brain itself. No ordinary radiograph could have shown it: a radiograph flattens the whole thickness of the body into one shadow, and soft tissues absorb X-rays so similarly that they all look the same grey. Hounsfield shared the 1979 Nobel Prize in Physiology or Medicine with Allan Cormack, who had worked out the mathematics independently. This lesson shows how CT, MRI and the endoscope each get round the limits of the plain radiograph that you met in Unit 7.
In a CT scanner an X-ray tube and an arc of several hundred detectors face each other across a ring, the gantry, and rotate together around the patient (Figure 4.1). Each detector measures the intensity that crosses the body along one ray. From Lesson 1, for a ray crossing small elements of thickness :
So each reading gives the sum of the attenuation coefficients along one line. One sum cannot say where along the line the absorber is, but sums measured along thousands of lines at hundreds of angles can. A computer solves for the value of in every small square (pixel) of the slice, using a method called filtered back-projection or newer iterative methods. Modern scanners move the patient steadily through the rotating gantry (a helical or spiral scan), collecting many slices in a few seconds, which the computer can also stack into a three-dimensional image.
Figure 4.1: A CT scanner viewed along the patient's body (not to scale).
Definition
Computed tomography (CT) is the reconstruction, by computer, of a cross-sectional image of the body from X-ray transmission measurements made along many lines at many angles. Each pixel shows a CT number in Hounsfield units (HU):
so water is 0 HU and air is about −1000 HU. Example: with for muscle and for water at 60 keV, muscle is .
Figure 4.2: The smallest possible CT problem: four unknown pixels and five measured ray sums.
Worked example 1: a four-pixel CT scan
A slice is divided into four pixels , each 1.0 cm wide (Figure 4.2). The measured values of are: row 1 () 0.81; row 2 () 0.40; column 1 () 0.40; column 2 () 0.81; diagonal () 0.42. (a) For row 1, what fraction of the X-rays got through? (b) Find for each pixel. (c) Using the CT numbers in the table below, identify each tissue.
Solution
(a) .
(b) Rows and columns alone give only four equations, and they are not independent (both pairs add to 1.21), so the diagonal is needed. From row 2 and column 1: , so . The diagonal gives , so . Then and .
(c) Comparing with the table: and (0.21) are muscle, (0.19) is fat and (0.60) is bone.
Lesson: projections from more directions are needed than there are unknowns would suggest. A real scanner solves for about 250 000 pixels per slice from about a million readings.
| Tissue | at 60 keV / | CT number / HU |
|---|---|---|
| air | 0.0002 | −999 |
| fat | 0.188 | −89 |
| water | 0.206 | 0 |
| muscle | 0.215 | +44 |
| cortical bone | 0.604 | +1935 |
Calculated by Ganzaa from NIST mass attenuation coefficients; real scanners use a spread of photon energies, so measured CT numbers differ somewhat.
Muscle and fat differ by only about 13% in , far too little to separate on a radiograph where many tissues overlap. In CT each pixel stands alone, so the 130 HU difference shows plainly; the display can even be set to stretch a narrow "window" of CT numbers over the whole grey scale. CT is fast and widely available, and it is the first test after a head injury or a stroke. Its cost is dose: from Lesson 2, a CT of the abdomen gives about 77 times the effective dose of a chest radiograph, so each CT scan must be justified.

Some parts of the body can be looked at directly with light, if the light can be taken round corners. An endoscope is a flexible tube passed through the mouth (to view the oesophagus, stomach and duodenum) or through a small cut (keyhole surgery). It relies on the optical fibre: a glass core of refractive index surrounded by a cladding of lower index . Light striking the core–cladding boundary at more than the critical angle , where , is totally internally reflected, and bounces along the fibre with very little loss, even when the fibre bends.
Figure 4.3: Light guided along an optical fibre by total internal reflection.
Light entering the end of the fibre is guided only if it arrives within the acceptance angle . Applying Snell's law at the end face (from air) and the critical-angle condition at the side gives
An endoscope contains two bundles. An incoherent bundle, with fibres arranged at random, carries light down to illuminate the tissue. A coherent bundle, with tens of thousands of fibres in exactly the same arrangement at both ends, carries the image back: each fibre carries one dot of the picture. Many modern endoscopes instead put a tiny camera chip at the tip and use the fibres only for lighting. Extra channels let the doctor blow in air, wash the lens, take a small tissue sample (biopsy), or pass a laser fibre to seal a bleeding vessel. Endoscopy uses visible light only, so it gives no ionising dose.
Worked example 2: designing a fibre
A fibre has a core of refractive index 1.52 and a cladding of 1.48. Find (a) the critical angle at the core–cladding boundary and (b) the acceptance angle in air. (c) Explain why the cladding is needed at all.
Solution
(a) , so .
(b) , so .
(c) Without cladding, total internal reflection would happen at a glass–air surface, and fibres touching each other in a bundle would let light leak from one to the next, scrambling the image. The cladding protects the reflecting surface and keeps each fibre's light separate.

MRI maps the hydrogen nuclei (protons) in the body's water and fat. A proton has spin, which makes it a tiny magnet. In the strong, steady field of the scanner (usually 1.5 T or 3.0 T, made by a superconducting coil cooled by liquid helium), each proton's magnetic axis precesses around the field direction, like a spinning top leaning in gravity, at the Larmor frequency:
Slightly more protons line up with the field than against it, so the tissue has a small net magnetisation along .
Figure 4.4: Precession, resonance and relaxation.
Definition
Magnetic resonance imaging (MRI) forms images from the radio signals emitted by hydrogen nuclei in a strong magnetic field after they absorb a radio-frequency pulse at their Larmor frequency, the frequency at which their magnetic moments precess about the field. Example: at 3.0 T, .
Worked example 3: the numbers behind MRI
For hydrogen in a 1.5 T scanner, find (a) the Larmor frequency, (b) the energy of one RF photon at this frequency, in eV, and compare it with Lesson 1, and (c) the fraction of excess protons aligned with the field at body temperature (310 K), using the approximation . (d) A gradient of 10 mT m⁻¹ is applied along the body. By how much do the Larmor frequencies of two slices 5.0 mm apart differ?
Solution
(a) , a radio frequency.
(b) , about 50 million times too small to ionise a water molecule (12.6 eV). MRI is non-ionising.
(c) : about 5 protons in every million. The signal is faint, which is why MRI needs strong fields, sensitive coils and long scans.
(d) , so . An RF pulse 2.1 kHz wide excites just one 5 mm slice.

MRI safety. There is no ionising radiation, but the magnet is always on. Iron and steel objects (scissors, oxygen cylinders, keys) are pulled violently into the bore, and some pacemakers and metal implants are unsafe, so every patient and visitor is screened before entering the room. The RF pulses warm the body slightly and the gradient noise is loud enough to require ear protection. Scans take longer than CT, and the narrow tunnel distresses some patients.
Warning
Common mistakes. (1) Saying MRI uses X-rays or "radiation" in the harmful sense: it uses a magnetic field and radio waves, which are non-ionising. (2) Saying MRI images the magnetic iron in blood: it images hydrogen nuclei (protons). (3) Thinking a CT image is a sharper radiograph: it is a computed map of in a slice, built from many projections. (4) Mixing up the bundles: the coherent bundle carries the image; the incoherent one only carries light in.
Tip
Link each machine to one equation. CT: (many sums give a map). Endoscope: (light trapped in glass). MRI: (frequency tells you where).
How does an MRI machine work?, by the Science Museum, London. Watch for the strong magnet aligning the protons, the radio pulse knocking them over and the signal given out as they realign.