Radiation and Medical Imaging
Senior 6 · Physics
Radiation and Medical Imaging
Lesson 1 of 5
25 min
A microwave oven running at 800 W pours about 50 kJ of energy into a cup of tea in one minute, and the tea simply gets hot. A dental X-ray delivers to the patient's jaw only a tiny fraction of a joule, yet the dentist steps out of the room and the patient may wear a lead collar over the thyroid gland. If total energy decided the danger, the oven would be the threat. It is not, because what matters for living cells is how the energy arrives: in how large a packet, and how densely along its path. In Units 5 and 7 you met alpha, beta and gamma radiation and X-rays. This lesson asks what any radiation actually does to the atoms of the body. The answer underlies every medical image in this unit, and every safety rule.
Medicine uses waves and particles across an enormous range of energies. They split into two families according to one question: can a single photon or particle knock an electron out of an atom or molecule?
Figure 1.1: The electromagnetic spectrum on a photon-energy scale. Only photons to the right of the dashed line carry enough energy to ionise the molecules of the body.
The energy needed to remove the outermost electron from the atoms and molecules of living tissue is about 10 eV to 15 eV; for a water molecule it is 12.6 eV. A photon carries , so photons can ionise only if : far ultraviolet, X-rays and gamma rays. Making the beam more intense adds more photons but does not make any one photon more energetic, which is why a powerful radio transmitter cannot ionise a single atom while a single 60 keV X-ray photon, through the electron it sets moving, can ionise nearly two thousand molecules.
Definition
Ionising radiation is radiation whose individual photons or particles carry enough energy to remove electrons from atoms or molecules, producing ions. Examples: X-rays, gamma rays, alpha and beta particles, neutrons, protons.
Non-ionising radiation is radiation whose photons or particles cannot ionise; it can still deposit energy, mainly as heat or by exciting molecules. Examples: radio waves, microwaves, infrared, visible light, most ultraviolet, and ultrasound (a mechanical wave, not an electromagnetic one).
Directly ionising radiation (charged particles: α, β, protons) ionises atoms along its track through the electric force. Indirectly ionising radiation (photons, neutrons) first gives its energy to charged particles, which then do the ionising.
| Radiation | Family | Medical use (this unit) | Main effect in tissue |
|---|---|---|---|
| radio-frequency waves | non-ionising | MRI (Lesson 4) | slight heating |
| visible and infrared light | non-ionising | endoscopy (Lesson 4) | heating, excitation |
| ultrasound | non-ionising (mechanical) | scanning (Lesson 3) | slight heating |
| X-rays | ionising, indirectly | radiography, CT, therapy | ionisation by secondary electrons |
| gamma rays | ionising, indirectly | nuclear medicine, therapy (Lesson 5) | ionisation by secondary electrons |
| β⁻ and β⁺ particles | ionising, directly | therapy; β⁺ in PET (Lesson 5) | ionisation along a track |
| α particles, protons | ionising, directly | targeted therapy | dense ionisation along a short track |
Ultraviolet sits on the border. UV-A and UV-B (wavelengths 280 nm to 400 nm) cannot ionise, but a UV-B photon (about 4 eV) can still excite DNA strongly enough to join neighbouring bases together. That is how sunlight causes sunburn and skin cancer without ionising anything: "non-ionising" does not mean "harmless".
Worked example 1: three photons compared with the threshold
Find the photon energy, in eV, of (a) a mobile-phone signal at 900 MHz, (b) green light of wavelength 550 nm, and (c) find the longest wavelength that can ionise a water molecule (ionisation energy 12.6 eV). Use , , .
Solution
(a) . In eV: , over three million times too small to ionise.
(b) . Enough to trigger the retina, not enough to ionise.
(c) , in the far ultraviolet.
Conclusion: the danger of a photon is set by its frequency, not by the power of the source.
A charged particle moving through matter pulls on the electrons it passes, ionising or exciting atom after atom until it has lost all its kinetic energy and stops. It therefore has a definite range. In air, about 34 eV is used up, on average, for every ion pair formed (this average is called the W-value; it is larger than the ionisation energy because some energy goes into excitation).
The energy a particle deposits per unit length of its track is its linear energy transfer (LET). Alpha particles in tissue have an LET of the order of 100 keV per micrometre; fast electrons, of the order of 0.2 keV per micrometre. High-LET radiation packs its damage so closely that both strands of a DNA molecule are often broken at once, which a cell finds much harder to repair. You will see this again in Lesson 2 as the radiation weighting factor.
Worked example 2: dense and sparse ionisation
Americium-241 emits alpha particles mainly at 5.49 MeV, which travel about 4.1 cm in air. A beta particle of 1.0 MeV travels a path of about 3.5 m in air. Taking W = 35 eV per ion pair for the alpha particle and 34 eV for the beta particle, find for each (a) the total number of ion pairs and (b) the average number per millimetre of path.
Solution
(a) Alpha: . Beta: .
(b) Alpha: per mm. Beta: per mm.
Interpretation: the two particles carry similar energies, but the alpha particle spends it about 450 times more densely. (Intermediate values were kept unrounded; the path lengths are typical values, so the answers are estimates to 2 significant figures.)
An X-ray or gamma photon has no charge, so it travels in a straight line until a single event removes it from the beam. Three events matter in medicine (Figure 1.2).
Figure 1.2: The three main interactions of X-ray and gamma photons with matter.
Worked example 3: why scattered photons blur an image
Technetium-99m, the most used tracer in nuclear medicine, emits gamma photons of 140 keV. One such photon is Compton-scattered through 90° inside the patient. Find its energy after scattering and the energy given to the electron. Take .
Solution
Dividing the Compton equation into energies (using ) gives .
With , : .
Electron: (subtracting the unrounded 109.9 keV).
Why it matters: the scattered photon leaves the body in a new direction, so if it were detected it would appear to come from the wrong place. Because it has lost about a fifth of its energy, the gamma camera of Lesson 5 can reject it by accepting only photons close to 140 keV.
Consider a narrow, parallel beam of photons, all of one energy, of intensity entering a thin layer of thickness . Each of the three processes removes photons at random, so the fraction removed is proportional to the thickness:
The constant is the linear attenuation coefficient (unit or ); it is the sum of the contributions of the three processes. Setting gives the half-value layer (HVL), the half-value thickness you used in Unit 5:
Because is proportional to the density, tables list the mass attenuation coefficient , in ; then . This has the same form as the decay law , with thickness in place of time.
| Material (density) | at 60 keV | at 1.25 MeV |
|---|---|---|
| skeletal muscle () | ||
| cortical bone () | ||
| water () |
Values from the NIST X-ray mass attenuation coefficient tables (Hubbell and Seltzer), rounded to 3 significant figures.
Conditions. The exponential law holds for a narrow beam of photons of a single energy. In a broad beam, some scattered photons still reach the detector, so the measured intensity is higher than . Charged particles do not obey it at all: they have a definite range.
Worked example 4: where contrast comes from
A narrow 60 keV X-ray beam crosses 6.0 cm of a limb along two paths: (A) 6.0 cm of muscle; (B) 4.0 cm of muscle and 2.0 cm of cortical bone. Use the table above. (a) Find the fraction transmitted along each path and their ratio. (b) Repeat for 1.25 MeV gamma rays and comment.
Solution
(a) ; (NIST values before rounding).
A: . B: . Ratio: .
(b) ; . A: . B: . Ratio: 1.10.
Comment: at 60 keV, photoelectric absorption by calcium makes bone stop about twice as many photons, so it stands out clearly. At 1.25 MeV, Compton scattering dominates and depends only on electron density, so bone and muscle look almost alike. Diagnostic X-ray tubes therefore run at about 25 kV to 150 kV, while radiotherapy uses MeV beams that pass through bone without being strongly absorbed by it.
This is a teacher demonstration with a sealed caesium-137 source (gamma rays of 662 keV), handled with tongs and returned to its lead container straight after use. A Geiger–Müller tube faces the source at a fixed distance, with a collimator so that the beam is narrow. The background count rate is measured first, with the source in its container, and subtracted from every reading. Lead sheets are then added one at a time.
| thickness x / mm | 0 | 2 | 4 | 6 | 8 | 10 | 12 |
|---|---|---|---|---|---|---|---|
| corrected count rate C / s⁻¹ | 846 | 661 | 503 | 395 | 299 | 236 | 178 |
| ln (C ÷ s⁻¹) | 6.74 | 6.49 | 6.22 | 5.98 | 5.70 | 5.46 | 5.18 |
Illustrative readings of the kind this demonstration gives.
Taking logarithms of gives : a straight line of gradient (Figure 1.3). A straight line confirms the exponential law; a curve that flattens at large would show scattered photons reaching the detector.
Figure 1.3: ln C against thickness of lead, with the line of best fit.

Worked example 5: analysing Experiment 1.1
Use the line of best fit in Figure 1.3 to find (a) the linear attenuation coefficient of lead for 662 keV gamma rays, (b) the half-value layer, and (c) the thickness of lead that would cut the count rate to 1% of its unshielded value.
Solution
(a) The line passes through (0, 6.75) and (12, 5.19). Gradient , so . (Published tables give about at this energy.)
(b) .
(c) , about 6.6 half-value layers.
Warning
Common mistakes. (1) Thinking that a stronger (more intense) source of light or radio waves becomes ionising: intensity adds photons; only frequency raises the energy per photon. (2) Writing for alpha or beta particles: charged particles have a range, not an exponential tail. (3) Forgetting to subtract the background count before taking logarithms; the error is largest for the thickest absorbers. (4) Mixing units: if is in , must be in cm.
Tip
Match each photon process to its job in medicine. Photoelectric: contrast (it depends strongly on , so it separates bone from flesh in a radiograph). Compton: dose and blur (it happens everywhere, deposits energy and scatters photons the wrong way). Pair production: high-energy therapy, and the positrons behind PET in Lesson 5.
Is radiation dangerous?, by Matt Anticole (TED-Ed). Watch for how the video separates ionising from non-ionising radiation by the energy of each packet, and compare its examples with Figure 1.1.
Practice
Key points
Sources and credits
The three diagrams, the experiment readings (illustrative), the worked examples and the practice exercises are written by Ganzaa. This lesson adapts CC BY-SA material and is shared under CC BY-SA 4.0.