In an antenatal clinic a midwife spreads cold gel on a pregnant woman's abdomen and presses a small probe against it. A grey, grainy picture appears on the screen, and moments later the room fills with a rapid whooshing beat: the baby's heart, beating more than twice as fast as the mother's. No radiation dose has been given, no needle used, and the examination can be repeated next month. In Unit 1 you met the basics: echoes from boundaries, the acoustic impedance and the reflected fraction . This lesson follows the pulse from the crystal that makes it to the picture on the screen, and asks what limits how deep and how sharp that picture can be.
Some crystals and ceramics, such as quartz, barium titanate and lead zirconate titanate (PZT), are piezoelectric: squeezing them produces a voltage across them, and a voltage across them makes them change thickness. An alternating voltage at the crystal's resonant frequency makes it vibrate strongly and send ultrasound into the body. When an echo returns, the same crystal is squeezed by it and produces a small voltage that the machine amplifies.
Definition
The piezoelectric effect is the production of a potential difference across certain crystals when they are mechanically stressed; the reverse piezoelectric effect is the change in their dimensions when a potential difference is applied. A transducer is a device that converts energy from one form to another; an ultrasound transducer converts electrical oscillations to ultrasound and back.
Figure 3.1: Inside an ultrasound transducer.
Worked example 1: matching the crystal to the body
Barium titanate has and ; soft tissue (muscle) has and . (a) Find both acoustic impedances. (b) What fraction of the intensity would be reflected straight back into a crystal pressed directly on tissue? (c) Find the impedance a matching layer should have so that the reflections at its two faces are equal, and the fraction reflected at each face.
Solution
(a) ; .
(b) . Four-fifths of the energy would never leave the probe.
(c) Equal reflections need , so . At each face .
Why it works: the layer is made a quarter of a wavelength thick. The wave reflected from its back face travels an extra half wavelength, so it cancels the equal wave reflected from its front face, exactly as in the anti-reflection coating on a lens. At the design frequency almost all the energy passes into the body.
The machine times each echo and assumes a speed of , the average for soft tissue, to place it at depth . The table recomputes the impedances from density and speed.
| Medium | / | / | / |
|---|---|---|---|
| air | 1.3 | 330 | 429 |
| water (similar to gel) | 1000 | 1500 | |
| blood | 1060 | 1570 | |
| fat | 925 | 1450 | |
| muscle (average) | 1075 | 1590 | |
| bone (varies) | 1400 to 1900 | 4080 | to |
Densities and speeds from OpenStax College Physics, Table 17.7.1; impedances recalculated by Ganzaa.
The intensity reflected at a boundary met head-on is and the intensity transmitted is . On its way down and back, a pulse also loses energy to absorption and scattering: in soft tissue the attenuation is roughly 0.5 dB per centimetre for each megahertz of frequency, where a loss of dB means the intensity is multiplied by . Deep echoes are therefore far weaker than shallow ones from the same kind of boundary, so the machine applies time-gain compensation (TGC): it amplifies each echo more the later it arrives.
Figure 3.2: The pulse-echo principle and the A-scan, before and after time-gain compensation.
Worked example 2: how strong is a deep echo?
A 5.0 MHz pulse passes from gel (treat as water) into 2.0 cm of fat, then 3.0 cm of muscle, and is reflected at a muscle–bone boundary (take the bone value ). Using the table, find (a) the fraction of the emitted intensity returning to the probe, ignoring attenuation, and (b) the fraction when attenuation of 0.5 dB per cm per MHz is included.
Solution
Gel to fat: , so .
Fat to muscle: , so .
Muscle to bone: .
(a) The pulse crosses each soft boundary twice (down and back): .
(b) The round trip is 10.0 cm: loss , a factor . Returning fraction .
Conclusion: attenuation, not reflection, is what makes deep echoes weak. That is why TGC is essential, and why bone, reflecting about 40% and absorbing strongly, leaves an "acoustic shadow" behind it.

Worked example 3: how many lines in a frame?
The scan in the photo shows tissue to a depth of 16.3 cm at 32 frames per second. Taking , find (a) the round-trip time to the deepest point, (b) the maximum pulse repetition frequency, and (c) the maximum number of scan lines in each frame.
Solution
(a) .
(b) .
(c) Lines per frame .
Trade-off: looking deeper or refreshing faster leaves fewer lines per frame, so the picture becomes coarser. Depth, frame rate and line density cannot all be increased together.
Higher frequency gives a shorter wavelength, , and shorter pulses, so finer detail; but attenuation grows in proportion to frequency, so the pulse cannot go as deep. At 3.5 MHz, and an echo from 12 cm deep (24 cm round trip) loses about . At 10 MHz, but the same echo would lose 120 dB, a factor of , and be lost in electronic noise. So abdominal and obstetric scans use about 2 MHz to 5 MHz, while the thyroid, tendons, the eye and blood vessels near the surface are scanned at 7 MHz to 15 MHz.
Echoes from moving red blood cells are shifted in frequency. In Unit 1 you derived for blood moving straight along the beam. In practice the beam meets the vessel at an angle (Figure 3.3), and only the velocity component along the beam counts:
Figure 3.3: Doppler geometry.
Colour Doppler shows flow towards the probe in one colour (usually red) and away in another (blue) on top of the B-scan; spectral Doppler plots the speeds against time, beat by beat. Doctors use it to find narrowed arteries, blood clots in leg veins and problems with blood flow to a baby through the umbilical cord. The angle must be kept below about 60°: as approaches 90°, approaches zero and small errors in the angle produce large errors in the speed.
Worked example 4: blood speed, and the danger of a wrong angle
A 5.0 MHz Doppler beam meets an artery at 60° to the flow and the measured shift is 1.30 kHz. (a) Find the blood speed (). (b) The true angle was 55°. What is the true speed, and what was the percentage error?
Solution
(a) .
(b) . Error .
A 5° error in the angle gave a 15% error in the speed, and the error grows rapidly at larger angles.

The ultrasonic distance sensors sold for school electronics and robotics projects work on exactly the same principle at 40 kHz. Point one at a wall 0.500 m away (measured with a metre rule), trigger a pulse from a microcontroller or a signal generator, and display the transmitted pulse and its echo on an oscilloscope. In one trial the echo arrived 2.92 ms after the pulse (illustrative reading), giving , close to the speed of sound in air at about 20 °C. Put a sheet of card halfway to the wall: two echoes appear, and the stronger one comes from whichever surface reflects more. That is an A-scan.
Ultrasound is not ionising, and no harmful effect has been shown at diagnostic intensities. It does deposit some energy as heat, and at high pressures it can make tiny gas bubbles oscillate (cavitation). Scanners therefore display a thermal index and a mechanical index, and operators keep the output and the scanning time as low as the examination allows: ALARA applies here too. Keepsake scans with no medical purpose are discouraged for this reason.

Warning
Common mistakes. (1) Forgetting the factor 2 in or in the Doppler shift: the sound goes there and back, and the moving blood shifts the frequency twice. (2) Using instead of : is the angle between the beam and the flow. (3) Forgetting that transmission at a boundary is , and that a pulse crosses each boundary twice. (4) Thinking a higher frequency is always better: it improves detail but costs depth.
Tip
For any ultrasound question, ask three things in order. Where? . How bright? at the boundary, at every boundary crossed, and the attenuation in dB. How fast? . Remember also: high frequency, shallow and sharp; low frequency, deep and coarse.
How to see with sound, by Jacques S. Abramowicz (TED-Ed). Watch how the piezoelectric probe both sends and listens, and why bone and air give strong echoes.
Practice
Key points
Sources and credits
The three diagrams, the experiment, the recalculated impedances, the worked examples and the practice exercises are written by Ganzaa.