Radiation and Medical Imaging
Senior 6 · Physics
Radiation and Medical Imaging
Lesson 5 of 5
25 min
Early each week, a busy nuclear-medicine department receives a heavy, lead-shielded container about the size of a small bucket. Staff call it the "cow". Inside, molybdenum-99 (half-life 65.9 h) is fixed to a small column of alumina, decaying into technetium-99m. Each morning a technologist runs salt water through the column and "milks" off the technetium, which has grown back overnight, then attaches it to chemicals that carry it to bone, heart muscle, the kidneys or the thyroid. After a week the molybdenum has decayed to about a sixth of its activity and the cow is replaced. Radiography, CT and ultrasound show the body's structure; nuclear medicine shows what its organs are doing. The same physics, turned up to much higher doses, is also used to destroy cancers.
A radiopharmaceutical is a radionuclide attached to a molecule that the body handles in a known way: a phosphate compound settles in growing bone, a glucose analogue collects in very active cells, iodine is taken up by the thyroid. Injected or swallowed, it goes where the chemistry takes it, and a detector outside the body records where the radiation comes from. The amounts are tiny (Worked example 1), so they do not disturb the processes they reveal.
A good imaging tracer:
| Radionuclide | Half-life | Emission used | Typical use |
|---|---|---|---|
| technetium-99m | 6.01 h | γ, 140 keV | bone, heart, kidney and thyroid scans |
| iodine-123 | 13.2 h | γ, 159 keV | thyroid imaging |
| fluorine-18 | 110 min | β⁺ (two 511 keV photons) | PET with FDG, a glucose analogue |
| iodine-131 | 8.02 days | β⁻ (and γ, 364 keV) | treating thyroid cancer and an overactive thyroid |
Technetium-99m is an excited state ("m" for metastable) of technetium-99; it decays to the ground state by emitting a gamma ray, with no beta particle, which keeps the patient's dose low.
Definition
A radioactive tracer (radiopharmaceutical) is a compound labelled with a radionuclide, given to a patient in a very small amount, so that its path and where it collects in the body can be followed from the radiation it emits. Example: technetium-99m attached to a phosphate compound collects where bone is being repaired or invaded by cancer.
Worked example 1: preparing a dose
A heart scan needs 740 MBq of technetium-99m at 11:00. (a) What activity must be prepared at 08:00? (b) How many technetium-99m nuclei are in the dose at 11:00, and what is their mass? (, molar mass , .)
Solution
(a) The dose decays for 3.0 h, so , about 1.05 GBq.
(b) ; . Mass , about 4 nanograms.
Figure 5.1: A gamma camera (not to scale).
Over several minutes the computer counts hundreds of thousands of photons into an image. If the camera rotates around the patient, the computer can reconstruct slices, as in CT: this is SPECT (single-photon emission computed tomography).

Worked example 2: inside the photomultiplier and the energy window
(a) A photomultiplier has 10 dynodes, and each releases 4 electrons for every electron that hits it. Find the number of electrons reaching the anode for each electron released by the photocathode. (b) The pulse-height analyser accepts energies within ±10% of 140 keV. Would a photon Compton-scattered through 30° be accepted? (.)
Solution
(a) Gain : one electron becomes about a million, a pulse big enough to measure.
(b) Window: 126 keV to 154 keV. . It is accepted. Photons scattered through small angles cannot be rejected, and they blur the image slightly: no energy window is perfect.

PET uses tracers labelled with positron emitters, above all fluorine-18 in FDG (fluorodeoxyglucose), which collects in cells that use a lot of glucose, such as many cancers, the working brain and heart muscle. Each positron travels about a millimetre in tissue and then annihilates with an electron, producing two 511 keV photons, as you saw in Unit 6. Because the pair had almost no momentum, the two photons fly off in opposite directions.
Figure 5.2: Coincidence detection in a PET scanner.
A ring of scintillation detectors surrounds the patient. When two detectors on opposite sides fire within a few nanoseconds of each other, a coincidence circuit records an annihilation somewhere on the straight line joining them, the line of response. No lead collimator is needed, because the pair of photons itself defines the direction, so PET is much more sensitive than a gamma camera. Millions of lines are reconstructed into slices. Modern scanners also measure the tiny difference in the arrival times (time of flight), which places the event at a distance from the midpoint of the line. Most PET scanners are combined with a CT scanner (PET/CT), so that the activity map can be laid over the anatomy; from Lesson 2, the combined effective dose is high, about 23 mSv for a whole-body scan.
Worked example 3: locating an annihilation by time of flight
In a PET scanner, the two 511 keV photons from one annihilation reach detectors A and B, which face each other across the ring, and the photon at B arrives 0.40 ns after the photon at A. (a) Where on the line did the annihilation happen? (b) Why must the coincidence window be only a few nanoseconds?
Solution
(a) If the event is a distance from the midpoint, towards A, the photon to B travels further. So and : 6.0 cm from the midpoint, on A's side.
(b) Light crosses a ring about 80 cm wide in under 3 ns, so two photons from one annihilation always arrive within a few nanoseconds. A longer window would accept more "random" pairs: two photons from different annihilations that happen to arrive close together, which put events on false lines.

Imaging tries to give the smallest useful dose. Radiotherapy does the opposite on purpose: it gives a tumour a dose high enough to kill its cells, typically a few tens of grays in total, while keeping the dose to the healthy tissue around it as low as possible. Every treatment plan is built around this balance.
Figure 5.3: (a) Crossfire from several beams. (b) Schematic depth-dose curves for megavoltage X-rays and for protons.
Megavoltage X-rays have two advantages. Compton scattering dominates, so bone does not absorb much more than soft tissue (Lesson 1, Worked example 4). And the fast secondary electrons are thrown forward, so the dose builds up over the first centimetre or two (about 1.5 cm for 6 MV) instead of being largest at the skin: the skin is spared. The total dose is usually given in daily fractions, for example 2 Gy a day, five days a week, because healthy cells repair radiation damage between fractions better than most tumour cells do.
Worked example 4: why several beams?
A tumour lies 10 cm deep. A cobalt-60 beam in tissue has . Ignoring the inverse-square law, scatter and build-up, find the dose where each beam enters the body if the tumour is to receive 2.0 Gy (a) from one beam, (b) from four equal beams from different directions, all 10 cm deep.
Solution
Fraction reaching the tumour: .
(a) Entrance dose , nearly twice the tumour dose.
(b) Each beam must deliver 0.50 Gy at the tumour: entrance dose .
Conclusion: with one beam the healthy tissue in front of the tumour receives more than the tumour; with four, the tumour receives 2.0 Gy and the tissue along each path at most about 0.94 Gy. Modern machines go further and spread the beam over a continuous arc.
| Method | What it shows | Ionising? | Main limitation |
|---|---|---|---|
| radiograph (Unit 7) | bone, lungs, foreign bodies | yes (low dose) | overlapping shadows; poor soft-tissue contrast |
| CT (Lesson 4) | cross-sections of all tissues, fast | yes (higher dose) | dose; less soft-tissue contrast than MRI |
| ultrasound (Lesson 3) | soft tissue, fluid, blood flow, the fetus, in real time | no | blocked by bone and gas; depends on the operator |
| MRI (Lesson 4) | soft tissue, brain, spinal cord, joints | no | slow, costly, noisy; unsafe with some metal implants |
| endoscopy (Lesson 4) | the inside surface of hollow organs, with biopsy | no | invasive; sees only surfaces |
| gamma camera, SPECT, PET | function: blood flow, metabolism, uptake | yes (from inside) | coarse detail; patient is radioactive for a while |
The best test answers the doctor's question with the least risk: justification and ALARA, from Lesson 2, applied to the choice of machine itself.
Warning
Common mistakes. (1) Saying the gamma camera sends radiation into the patient: the radiation comes out of the patient, from the tracer. (2) Choosing an alpha or beta emitter for imaging: those particles cannot reach a detector outside the body. (3) Saying the collimator focuses gamma rays like a lens: it cannot bend them; it only absorbs those not travelling along its holes. (4) Thinking a radiotherapy patient becomes radioactive: external beams leave no activity behind; only patients given a source (brachytherapy or iodine-131) carry activity for a while.
Tip
For any tracer question check "G-E-H-C": Gamma emitter? Energy about 100 keV to 200 keV? Half-life of hours? Chemistry that takes it to the organ? For therapy, reverse the first: choose a beta (or alpha) emitter, because you want the energy to stay inside.
How does a PET scan work?, by Imperial College London. Watch for how the tracer is made and injected, and how pairs of photons reaching opposite detectors build the image.
Practice
Key points
Further reading
Sources and credits
The three diagrams, the worked examples and the practice exercises are written by Ganzaa. This lesson adapts CC BY-SA material and is shared under CC BY-SA 4.0.